DUPLEX PLATE PHANTOM
Article No.: 11 0017X
Heavy Metal – HM
Light Metal – LM
according to ASTM E2737
Duplex Plate Phantom Type 2
for Perfomance Evaluation
and Long-Term-Stability Tests of DDAs
Test Phantom for Long-Term Stability Tests of DDA Systems (DR)
The Type 2 Duplex Plate Phantom enables to test relevant parameters of DDA systems like Basic Spatial Resolution, Contrast Sensitivity, SNR and Signal Level for initial performance evaluation and long-term stability. These test required and described in detail in ASTM E2737 have to be performed periodically for checking process control and lang-term stability.
IQI, Gauge | Test, Check | |
A | Base Plate – thin | HM 5 mm / LM 12 mm |
B | Top Plate – thick | HM 15 mm / LM 38 mm |
C | Hole Type IQI – .SS 25 / AL .50 | Contrast Sensitivity (edge or hole) |
D | Hole Type IQI – SS .81 / AL 2.0 | Contrast Sensitivity (edge or hole) |
E | ROIs – Thin Plate | Signal to Noise Ratio, Signal Level |
F | ROIs – Thick Plate | Signal to Noise Ratio, Signal Level |
G | Duplex Wire Type IQI | Basic Spatial Resolution, Unsharpness |
Dimensions: | 400 x 300 x 15 / 38 mm | |
15.75 x 11.81 x 0.59 / 1.50 inch |
Types:
Article No.:
Duplex Plate Phantom Type 2 Heavy Metal (HM) in wooden case
Complete with Hole Type IQI ASTM E1742 – SS size .25 and .81
Test Certificates, Declaration of Conformity
11 00177
Duplex Plate Phantom Type 2 Light Metal (LM) in wooden case
Complete with Hole Type IQI ASTM E1742 – AL size .50 and 2.0
Test Certificates, Declaration of Conformity
11 00178
Optional:
Holder for Duplex Plate Phantom Type 2, stainless steel
for clamping and easy positioning of both versions – HM and LM
11 00179
Duplex Wire Type IQI acc. to ASTM E2002, ISO 13232-5 – D13
Duplex Wire Type IQIs not included but required.
11 00155
You can find accessories on the leaflet
leaflet German:
leaflet English: DUPLEX PLATE PHANTOM